Journal Information
Journal ID (publisher-id): BM
Journal ID (nlm-ta): Biochem Med (Zagreb)
Title: Biochemia Medica
Abbreviated Title: Biochem. Med. (Zagreb)
ISSN (print): 1330-0962
ISSN (electronic): 1846-7482
Publisher: Croatian Society of Medical Biochemistry and Laboratory Medicine
Article Information
Copyright statement: ©Croatian Society of Medical Biochemistry and Laboratory Medicine.
Copyright: 2022, Croatian Society of Medical Biochemistry
License (open-access):
This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Date received: 17 January 2022
Date accepted: 28 March 2022
Publication date (print and electronic): 15 June 2022
Volume: 32
Issue: 2
Electronic Location Identifier: 020710
Publisher ID: bm-32-2-020710
DOI: 10.11613/BM.2022.020710
Verification of a 6-part differential haematology analyser Siemens Advia 2120i
Introduction
The aim of this study was to perform a comprehensive verification of a 6-part differential haematology analyser Siemens Advia 2120i (Erlangen, Germany), prior to its routine implementation.
Materials and methods
Our verification protocol included: precision (within- and between-run), estimated bias (%) as measure of trueness, which was calculated from observed and manufacturers’ declared value, analytical measuring interval (AMI), carryover, confirmation of a limit of blank (LoB), determination of a limit of detection (LoD) and limit of quantitation (LoQ). The K2 ethylenediaminetetraacetic acid (EDTA) patients’ leftover samples were used for verification of analyser Advia 2021i. Acceptance criteria were based on manufacturer technical specifications (Siemens), 2016 state-of-the-art criteria (Vis and Huisman), and EFLM Biological Variation Database.
Results
The within- and between-run precision were acceptable for all parameters and the lowest coefficients of variation were observed for mean corpuscular volume (MCV) (0.3% and 0.6%, respectively). Estimated bias was within the acceptance criteria for all parameters except for MCV (the estimated bias was 2.2% (acceptance criteria 2.0%). AMI was confirmed for all tested parameters (r > 0.99). The carryover estimates ranged from 0.1% for platelet count (Plt) to 0.6% for red blood cell count and were within the manufacturers’ specifications (≤ 1%). Manufacturers’ claims for LoB were confirmed for leukocytes, erythrocytes, haemoglobin, and platelets. The estimated LoD and LoQ were 0.05 x109/L and 0.1 x109/L for white blood cell count, while for Plt values were 2 x109/L and 3 x109/L, respectively.
Keywords: verification; haematology analyser; haematology; blood cell count