Journal Information
Journal ID (publisher-id): BM
Journal ID (nlm-ta): Biochem Med (Zagreb)
Title: Biochemia Medica
Abbreviated Title: Biochem. Med. (Zagreb)
ISSN (print): 1330-0962
ISSN (electronic): 1846-7482
Publisher: Croatian Society of Medical Biochemistry and Laboratory Medicine
Article Information
Copyright statement: Croatian Society of Medical Biochemistry and Laboratory Medicine.
Copyright: 2017, Croatian Society of Medical Biochemistry
License (open-access):
This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Date received: 31 July 2017
Date accepted: 24 April 2018
Publication date (print and electronic): 15 June 2018
Volume: 28
Issue: 2
Electronic Location Identifier: 020712
Publisher ID: bm-28-2-020712
DOI: 10.11613/BM.2018.020712
Evaluation of the analytical performances of Cobas 6500 and Sysmex UN series automated urinalysis systems with manual microscopic particle counting
Introduction
Automated urinalysis systems are valuable tools in clinical laboratories, especially those with a high work load. The objective of this study was to compare the analytical performance of Sysmex UN series urine analyser, which may become a new one in our laboratory, with the Cobas 6500 automated urine analyser, which is used in our laboratory for a long time. For comparisons, manual microscopical examination was accepted as reference method.
Materials and methods
A total of 470 urine samples were tested in the two automated urinalysis systems, and urine sediment testing with manual microscopy was applied to a 100 pathological samples of the total 470. The diagnostic performance of the two automated urine analysers was compared with each other and manual microscopy.
Results
Differences were determined between automated and manual microscopy in some pathological samples. The resultant regression equations were as follows. Comparison of Cobas U701 with Sysmex UF-5000: y = - 0.57 (- 0.85 to - 0.29) + 0.95 (0.92 to 0.99) x for RBC, and y = - 0.11 (- 0.54 to 0.29) + 0.89 (0.84 to 0.98) x for WBC; comparison of Cobas U701 with manual microscopy: y = - 0.45 (- 0.85 to 0.21) + 1.00 (0.92 to 1.07) x for WBC; and comparison of Sysmex UF-5000 with manual microscopy: y = - 0.74 (- 1.09 to - 0.57) + 0.87 (0.85 to 0.91) x for WBC.
Keywords: urinalysis; automated urinalysis; manual microscopy