Journal Information
Journal ID (publisher-id): BM
Journal ID (nlm-ta): Biochem Med (Zagreb)
Title: Biochemia Medica
Abbreviated Title: Biochem. Med. (Zagreb)
ISSN (print): 1330-0962
ISSN (electronic): 1846-7482
Publisher: Croatian Society of Medical Biochemistry and Laboratory Medicine
Article Information
Copyright statement: ©Croatian Society of Medical Biochemistry and Laboratory Medicine.
Copyright: 2020, Croatian Society of Medical Biochemistry
License (open-access):
This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Date received: 20 July 2019
Date accepted: 03 December 2019
Publication date (print and electronic): 15 February 2020
Volume: 30
Issue: 1
Electronic Location Identifier: 010706
Publisher ID: bm-30-1-010706
DOI: 10.11613/BM.2020.010706
Analytical verification and comparison of chromogenic assays for dabigatran, rivaroxaban and apixaban determination on BCSXP and STA Compact Max analysers
Introduction
The aim of the study was to perform analytical verification and comparison of chromogenic assays for determination of dabigatran, rivaroxaban and apixaban concentration on BCSXP and STA Compact Max analysers.
Materials and methods
Precision, linearity, measurement uncertainty estimation and determination of limit of blank, limit of determination and limit of quantification were calculated. Analytical performance specifications were set according to manufacturer specifications and literature data on between laboratory variability. Comparison of the methods was done using Bland-Altman and Passing-Bablok regression analysis.
Results
Obtained results have shown acceptable precision on STA Compact Max only for dabigatran (CV = 3.5%) at lower concentration level comparing to manufacturer declaration (CV = 3.6%). On BCSXP, the highest coefficient of variation has been shown for apixaban (6.1%) at lower concentration level. Within laboratory precision was not met on STA Compact Max for all assays. Bland-Altman analysis has shown statistically significant bias for dabigatran (23.2%, 95%CI 11.2 – 35.3; P < 0.001) and apixaban (8.4%, 95%CI 1.2 – 15.6; P = 0.023). Passing-Bablok regression analysis has shown systematic and proportional deviation between methods for rivaroxaban (y = 6.52 (2.94 to 11.83) + 0.84 (0.80 to 0.89) x.
Conclusion
Chromogenic assays for dabigatran, rivaroxaban and apixaban on BCSXP and STA Compact Max analysers are shown as methods with satisfactory long-term analytical performance specifications for determination of direct oral anticoagulants in clinical laboratories. However, we cannot recommend interchangeable use because of the significant bias between assays.
Keywords: chromogenic assay; dabigatran; rivaroxaban; apixaban; method comparison