Journal Information
Journal ID (publisher-id): BM
Journal ID (nlm-ta): Biochem Med (Zagreb)
Title: Biochemia Medica
Abbreviated Title: Biochem. Med. (Zagreb)
ISSN (print): 1330-0962
ISSN (electronic): 1846-7482
Publisher: Croatian Society of Medical Biochemistry and Laboratory Medicine
Article Information
Copyright statement: ©Croatian Society of Medical Biochemistry and Laboratory Medicine.
Copyright: 2020, Croatian Society of Medical Biochemistry
License (open-access):
This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Date received: 26 February 2020
Date accepted: 26 May 2020
Publication date (electronic): 05 August 2020
Publication date (print): 15 October 2020
Volume: 30
Issue: 3
Electronic Location Identifier: 030701
Publisher ID: bm-30-3-030701
DOI: 10.11613/BM.2020.030701
Analytical bias of automated immunoassays for six serum steroid hormones assessed by LC-MS/MS
Author notes:
[*] Corresponding author: i.markov00@gmail.com
Introduction
There is a growing amount of evidence showing the significant analytical bias of steroid hormone immunoassays, but large number of available immunoassays makes conduction of a single comprehensive study of this issue hardly feasible. Aim of this study was to assess the analytical bias of six heterogeneous immunoassays for serum aldosterone, cortisol, dehydroepiandrosterone sulphate (DHEAS), testosterone, 17-hydroxyprogesterone (OHP) and progesterone using the liquid chromatography coupled to the tandem mass spectrometry (LC-MS/MS).
Materials and methods
This method comparison study included 49 serum samples. Testosterone, DHEAS, progesterone and cortisol immunoassays were performed on the Abbott Architect i2000SR or Alinity i analysers (Abbott Diagnostics, Chicago, USA). DiaSorin’s Liaison (DiaSorin, Saluggia, Italy) and DIAsource’s ETI-Max 3000 analysers (DIAsource ImmunoAssays, Louvain-La-Neuve, Belgium) were chosen for aldosterone and OHP immunoassay testing, respectively. All immunoassays were evaluated against the LC-MS/MS assay relying on the commercial kit (Chromsystems, Gräfelfing, Germany) and LCMS-8050 analyser (Shimadzu, Kyoto, Japan). Analytical biases were calculated and method comparison was conducted using weighted Deming regression analysis.
Keywords: steroid hormone; immunoassay; bias; LC-MS/MS; method comparison